Subcrestal Implant
Placement Protocol
Certain® PREVAIL® 4/3mm, OSSEOTITE® Certain And
OSSEOTITE External Hex Implant Placement
(continued from previous
page)
11.
Pick up the implant from the
surgical tray using the Certain® Implant Placement Driver Tip
(IIPDTS or IIPDTL).
NOTE: The Certain MicroMiniplant™ (3.25mm dia.)
requires the use of a dedicated MicroMiniplant Driver Tip
(IMPDTS or IMPDTL) that is marked with a purple band on the
shank. The Internal Hex configuration of the MicroMiniplant is
smaller than the standard Certain Internal Hex Connection (4,
5 and 6mm dia. implants). The item numbers can be identified
on the side of the driver tip.
or
Pick up the implant from the surgical tray
using the Handpiece Connector (MDR10).
Carry the implant to the mouth facing upward to
prevent accidental dislodging.
12.
Place the implant in the prepared site at
approximately 15–20rpm. The ICE® Design will allow smooth and
precise implant placement without tapping in all but the most
dense bone. The recommended drill torque setting for placing
an implant is 40–50Ncm.
13.
To remove the Certain Implant Placement Driver
Tip from the implant, lift straight up and out.
or
To remove the implant mount, place the Open- End Wrench
onto the mount. Loosen the screw at the top of the mount with
a Large Hex Driver or the Large Hex Driver Tip inserted into
the Right-Angle Driver and rotate counter-clockwise. After the
screw is loosened, rotate the Open-End Wrench
counter-clockwise slightly before removing the mount. The
mount may be carried from the mouth with the Open-End
Wrench.
- Instruments needed:
Open-End Wrench (CW100), Large
Hex Driver Tip (RASH3) and Right-Angle Driver (CATDB with
CADD1) or a Large Hex Driver (PHD02N)
14.
Pick up the cover screw from the surgical tray using
the Certain Implant Placement Driver Tip (IIPDTS or IIPDTL) or
Large Hex Driver (PHD02N) and place onto the implant.
NOTE: When using the Certain Implant Placement Driver,
reduce the torque setting on the drilling unit to 10Ncm.
or
Pick up the cover screw from the surgical tray
using the Small Hex Driver (PHD00N) and place onto the
implant. The Small Hex Driver may be used for all External Hex
Cover Screws. Final hand-tightening of the cover screw should
be done with the Small Hex Driver.
NOTE: At this step, a temporary healing abutment may be
placed for single-stage surgery instead of a cover screw.
15.
Close tissue and suture.