Subcrestal Implant Placement Protocol In Soft Type IV
Bone OSSEOTITE®
Tapered Certain® And OSSEOTITE Tapered External Hex
Implant Placement (continued
from previous
page)
12. Pick up the implant from the
surgical tray using the Certain Implant Placement Driver Tip.
NOTE: The Certain MicroMiniplant™ (3.25mm dia.)
requires the use of a dedicated MicroMiniplant Driver Tip
(IMPDTS or IMPDTL) that is marked with a purple band on the
shank. The Internal Hex configuration of the MicroMiniplant is
smaller than the standard Certain Internal Hex Connection (4,
5 and 6mm dia. implants). The item numbers can be identified
on the side of the driver tip.
or
Pick up the implant from the surgical tray
using the Handpiece Connector.
Carry the implant to the mouth facing upward to
prevent accidental dislodging.
Instruments needed: Implant Placement Driver Tip
(IIPDTS or IIPDTL) Handpiece Connector (MDR10)
13. Place the implant in the prepared site at
approximately 15–20rpm. It is not uncommon for the handpiece
to stall before the implant is completely seated. The implant
position must match what was simulated with the NTDI or there
is a risk of a poor fit between the implant and osteotomy. In
dense bone, it is recommended to tap the site prior to implant
placement.
14. Final seating of the implant may require the
use of the Ratchet Extension and the Ratchet Wrench.
Instruments needed: Ratchet Wrench (WR150)
Certain Ratchet Extension (IRE100 or IRE200)
External Hex Ratchet Extension (RE100 or RE200)