Subcrestal Implant Placement Protocol In Soft Type IV Bone 
OSSEOTITE® Tapered Certain® And OSSEOTITE Tapered External Hex Implant Placement
(continued from previous page)

15. undefined To remove the Certain Implant Placement Driver Tip from the implant, lift straight up and out.

undefined To remove the implant mount, place the Open-End Wrench onto the mount. Loosen the screw at the top of the mount with a Large Hex Driver or the Large Hex Driver Tip inserted into the Right-Angle Driver and rotate counter-clockwise. After the screw is loosened, rotate the Open-End Wrench counter-clockwise slightly before removing the mount. The mount may be carried from the mouth with the Open-End Wrench.

  • Instruments needed:
    Open-End Wrench (CW100), Large Hex Driver Tip (RASH3) and Right-Angle Driver (CATDB with CADD1) or a Large Hex Driver (PHD02N)

16. undefined Pick up the Cover Screw from the No-Touch™ Implant Tray with the Implant Driver or Large Hex Driver and place onto the implant.

NOTE: When using the Certain Implant Placement Driver, reduce the torque setting on the drilling unit to 10Ncm.

  • Instruments needed:
    Implant Placement Driver Tip (IIPDTS or IIPDTL) Large Hex Driver (PHD02N)

17. undefined Pick up the Cover Screw from the No-Touch Implant Tray with the Small Hex Driver and place onto the implant. 

  • Instruments needed:
    Small Hex Driver (PHD00N)

NOTE: At this step, a temporary healing abutment may be placed for single-stage surgery instead of a cover screw.

18. undefined Close the tissue and suture.

undefined

undefined

 

 

 

 

 
 

BIOMET 3i | All Rights Reserved