Implant Level Impressions

Choices
Pick-Up Impression Coping
The Pick-Up Impression Coping transfers the position of an implant hex to a laboratory cast. The coping is picked up inside the impression when the impression is removed from the mouth. An open top impression tray is used with the Pick-Up Impression Coping.

Note
• Pick-Up Impression Copings may be difficult to use in limited, interarch distance spaces.
• Pick-Up Impression Copings are ideal for use when an impression is made of multiple divergent implants.

Twist Lock™ Transfer Coping
The Twist Lock Transfer Coping transfers the position of an implant hex to the laboratory cast. The coping is designed to stay on the implant in the mouth when the impression is removed. The coping is then removed from the implant and transferred into the impression. A closed top impression tray is used with the Twist Lock Transfer Coping.

Note
• Twist Lock Impression Copings are ideal for use in limited, interarch distance spaces.
• Twist Lock Impression Copings are contraindicated for use when an impression is made of multiple divergent implants. There is a risk of not being able to remove the impression tray from the copings once the impression material has set.

Selecting The Proper Impression Coping
Select the proper platform diameter Pick-Up or Twist Lock Transfer Impression Coping(s) by matching the color of the implant platform(s). Also, match the diameter of the coping to the EP® Diameter of the healing abutment. See healing abutment measurements:

Healing Abutment Measurements:

undefined

If using the Pick-Up Impression Coping, turn to page 11.
If using the Twist Lock Transfer Impression Coping, turn to page 14.

Components Needed
• Matching Platform and EP Diameter Pick-Up or Twist Lock Impression Coping(s).
• Matching implant laboratory analog(s).

Instruments Needed
• PHD02N or PHD03N - .048in. Large Hex Driver
• ICD00 - Impression Coping Driver (For Twist Lock Only).

 

 

 

 

 