Implant And Abutment
Level Impressions Pick-Up Impression
Copings
1.
Restorative Dentist Select the proper Pick-Up Impression Coping by
matching the EP® Diameter of the healing abutment and matching
the color of the implant platform to determine platform
diameter (see below). Remove the healing abutment from the
implant using the Large Hex Driver. To help prevent accidental
swallowing, thread floss through the spinner on the driver.
2. Activate the fingers using the QuickSeat® Activator
Tool. Place the Pick-Up Impression Coping into the implant,
line up the hex and press firmly until feeling the tactile
click.
or
Place the Pick-Up Impression Coping on the
implant and engage the hex.
Thread the Pick-Up Impression Coping Screw into
the implant until finger tight. Tighten the screw using the
Large Hex Driver.
3. Radiograph the interface to verify complete seating
of the coping on the implant. Place the film perpendicular to
the interface of the coping on the implant or abutment.
4. A custom or stock open top
impression tray is used for the Pick-Up Impression Technique.
Cut a small hole in the tray through which the screw may
protrude. Try in the tray to verify that the screw head is
visible through the opening.