Implant And Abutment
Level Impressions Twist Lock™ Transfer Impression
Copings (continued from previous
page)
9. Immediately replace the healing
abutment on the implant using the Large Hex Driver.
10.
Laboratory Place the proper diameter Implant Lab Analog
into the impression coping engaging the hex. Hold the
components together while finger tightening the screw. Verify
that the impression coping is seated passively on the analog.
11. Re-index the impression
coping/analog assembly into the impression using firm pressure
to its full depth. Slightly rotate the coping/analog clockwise
until feeling antirotational resistance. This indicates that
the orientation grooves are locked into place and the implant
hex is accurately transferred.
12. Syringe a soft-tissue material
around the coping and analog interface. Pour the cast in die
stone. Articulate with the opposing cast.