Provide® Abutment Placement
(continued from previous page)

4. Place the Provide Abutment into the implant. Line up the hexes and place the flat side of the post to the buccal. Press firmly until hearing and feeling the audible and tactile click.

5. Thread the Certain® Gold-Tite® Abutment Screw into the implant until finger tight using a .048" Large Hex Driver. Radiograph the interface to verify an accurate fit. Torque the abutment screw to 20Ncm using a .048" Large Hex Driver Tip and a torque device. Seal the top of the access hole with temporary filling material.