Screw Retained Single-Unit Provisional PreFormance® Temporary Cylinder Indirect Technique

Material:
undefined PEEK; titanium alloy connection

Indications:
• Screw Retained Single-Unit Provisional Restorations
• Minimum Interarch Space of 4mm
• Minimum Tissue Height of 2mm

1. Restorative Dentist
undefined Follow the steps for Implant Level Impressions on pages 11-13 for the Pick-Up Technique and pages 14-16 for the Twist Lock™ Transfer Technique.

2. Laboratory
undefined Set a denture tooth in wax on the cast where the single tooth is missing.

3. undefined Make a vacuum formed template over the denture tooth and adjacent teeth on the cast. Remove the template, denture tooth and wax from the cast.

4. undefined Select the proper diameter Hexed PreFormance Temporary Cylinder by matching the color of the implant platform. Place it into the implant analog, line up the hex and press firmly until hearing and feeling the audible and tactile click. (See page 6 for finger deactivation instructions.)

or

undefined Place the proper diameter Hexed PreFormance Temporary Cylinder onto the implant analog and engage the hex.

undefined Thread a waxing screw into the analog until finger tight using a .048” Large Hex Driver.

undefined

undefined

(continued on next page)

 

 

 

 

 

 
 