Screw Retained Single-Unit
Conical Abutment
Indirect Technique
(continued from previous page)

5. undefined Torque the Conical Abutment Screw into the implant at 20Ncm using the Abutment Driver Tip and a torque device. (If the surgeon places the abutment, a Conical Abutment Healing Cap is threaded onto the abutment using a .048” Large Hex Driver.)

6. Restorative Dentist
undefined Follow the steps for Abutment Level Impressions on page 11-13 for the Pick-Up Impression Technique and pages 14-16 for the Twist Lock™ Transfer Technique.

7. Laboratory
undefined Place a Hexed Conical Gold or Castable Cylinder onto the conical analog. Thread with a waxing screw or retaining screw into the analog until finger tight using a .048” Large Hex Driver. Reduce or adjust the plastic sleeve as necessary. Wax the single-unit PFM coping to the waxing sleeve.

8. undefined Remove the retaining or waxing screw and carefully remove the wax coping from the implant analog. Invest, burnout and cast the single-unit PFM coping to the Conical Gold Cylinder using a semi-precious or high noble alloy (see page 9 for casting alloy specifications). Divest and finish the coping.

undefined

undefined

(continued on next page)

 

 

 

 

 
 