5. Torque the Conical Abutment Screw into the
implant at 20Ncm using the Abutment Driver Tip and a torque
device. (If the surgeon places the abutment, a Conical
Abutment Healing Cap is threaded onto the abutment using a
.048” Large Hex Driver.)
6.
Restorative Dentist Follow the steps for Abutment Level Impressions
on page 11-13 for the Pick-Up Impression Technique and pages
14-16 for the Twist Lock™ Transfer Technique.
7.
Laboratory Place a Hexed Conical Gold or
Castable Cylinder onto the conical analog. Thread with a
waxing screw or retaining screw into the analog until finger
tight using a .048” Large Hex Driver. Reduce or adjust the
plastic sleeve as necessary. Wax the single-unit PFM coping to
the waxing sleeve.
8. Remove the retaining or waxing screw and
carefully remove the wax coping from the implant analog.
Invest, burnout and cast the single-unit PFM coping to the
Conical Gold Cylinder using a semi-precious or high noble
alloy (see page 9 for casting alloy specifications). Divest
and finish the coping.