Bar Supported
Overdenture
Standard Abutment
Hader
Bar
Material:
Titanium Alloy
Commercially Pure Titanium
Indications:
• Multiple
Implant, Bar Retained and Fixed Removable Overdentures
•
Minimum Space For Bar Construction is 7mm
• Maximum
Divergence of 30°
1. Surgeon Or Restorative
Dentist
Select the proper Standard Abutment collar
height for each implant by measuring the height of the healing
abutment above tissue, not including the dome portion.
Subtract the measurement from the total height of the healing
abutment and add 1mm. Match the color of the implant platform.
2.
Remove the healing abutments from the implants,
one at a time, using a .048” Large Hex Driver. To help prevent
accidental swallowing, thread floss through the spinner on the
driver.
3.
Place the proper height Standard Abutment into
the implant and thread into place until finger tight.
or
Place the proper height Standard Abutment
collar onto the implant, engaging the hex and thread the
Standard Abutment Screw into the implant by turning the
spindle on the ASYST® Tool until finger tight. Remove the
ASYST Tool.
4.
Radiograph the interfaces to verify complete
seating of the abutments on the implants. Place the film
perpendicular to the interface of the abutment on the
implant.