Bar Supported Overdenture
Standard Abutment
Hader Bar

Material:
undefined Titanium Alloy
undefined Commercially Pure Titanium

Indications:
• Multiple Implant, Bar Retained and Fixed Removable Overdentures
• Minimum Space For Bar Construction is 7mm
• Maximum Divergence of 30°

1. Surgeon Or Restorative Dentist
undefined Select the proper Standard Abutment collar height for each implant by measuring the height of the healing abutment above tissue, not including the dome portion. Subtract the measurement from the total height of the healing abutment and add 1mm. Match the color of the implant platform.

2. undefined Remove the healing abutments from the implants, one at a time, using a .048” Large Hex Driver. To help prevent accidental swallowing, thread floss through the spinner on the driver.

3. undefined Place the proper height Standard Abutment into the implant and thread into place until finger tight.

or

undefined Place the proper height Standard Abutment collar onto the implant, engaging the hex and thread the Standard Abutment Screw into the implant by turning the spindle on the ASYST® Tool until finger tight. Remove the ASYST Tool.

4. undefined Radiograph the interfaces to verify complete seating of the abutments on the implants. Place the film perpendicular to the interface of the abutment on the implant.