Indications: • Tissue
Supported Removable Overdentures on Two to Four Implants •
Partially Edentulous Overdentures With One or More Implants
• Minimum Interarch Distance of 4mm • Angle Correction
With Up to 40º Between Divergent Implants
1. Surgeon Or Restorative
Dentist Select the proper LOCATOR Abutment collar
height for each implant by measuring the height of the healing
abutment above tissue, not including the dome portion.
Subtract the measurement from the total height of the healing
abutment and add 1mm.
2. Remove the healing abutments from the implants,
one at a time, using a .048” Large Hex Driver. To help prevent
accidental swallowing, thread floss through the spinner on the
driver.
3. Thread the proper height LOCATOR Abutment into
the implant using the plastic delivery tool packaged with the
abutment. Use the LOCATOR Driver to finger tighten.
4. Radiograph the interfaces to verify complete
seating of the abutments on the implants. Place the film
perpendicular to the interface of the abutment on the
implant.